2IML-S3Q04 Gas Chromatograph

● C1 ~ C5 analysis cycle can be switched between 30s / 90s (with backflush)
● Embedded workstation to improve the speed and reliability of signal acquisition
● Full intelligent control, automatic ignition, automatic zero calibration, automatic fault diagnosis
● Large screen display, touch screen operation
● With serial, parallel, network, recorder and other output interfaces
● Data storage for months, not only for online use, but also for independent use
● Power supply: 220V ± 10% AC, 50 ± 1 Hz
● Total power: ≤800 W
● Working temperature: 0 ℃ ~ 35 ℃
● Relative humidity: 45% to 85%
● Analysis cycle: 30 / 90s switchable (with backflush)
● Methane ethane resolution: 100: 1 C1 / C2 clearly distinguishable, resolution ≥ 0.95
● Minimum detectable concentration: total hydrocarbons: 10 μmol / mol
Composition: 5μmol / mol
● Measurement range: Total hydrocarbons: 10 ~ 1 × 106 μmol / mol
Components: 5 ~ 1 × 106 μmol / mol
● Noise: ≤0.5% F.S
● Baseline drift: ≤1% F.S./h
● Repeatability error: ≤2.5% F.S.